Brunswick Biomedical

Heart-Lung Resuscitator

HLR

  • Performs heart-lung resuscitation automatically.
  • Oxygen powered for complete portability-no electrical components are used.
  • Contoured backboard hyper-extends the patient’s neck, which aids in maintaining an open airway.
  • Consistent and uniform number, depth, and rate of chest compression’s interposed with volume cycled ventilation (factory set at 90 chest compression’s/minute-ventilation’s of 100% oxygen after every fifth compression).
  • The Quick-Fit chest compressor can be applied in five seconds.
  • Two trained people can set up and operate; once the unit is in operation, only one person is needed for monitoring.
HLR Features

Chest Straps-adjust instantly for proper fit.

Shoulder and Groin Straps- help prevent patient movement during transportation and resuscitation.

Ventilation Volume Control-tailors ventilation for each individual patient.

Ventilation Hose-reinforced to prevent kinking and occlusion. A built-in standard 15-mm fitting permits direct connection of ventilation hose to endotracheal tube, esophageal airway or face mask.

Relief Valve-pops off at 60 cm H2O mask pressure maximum-prevents excessive pressure.

Quick-Connect Dual Oxygen Inputs-permits continuous operation when changing oxygen sources.

Chest Compressor Thrust Control-permits precise chest compression for each individual patient.

Travels with Patient- low profile compressor allows for transport in and out of any rescue vehicle without disturbing patient.

Fully insulated- allows the physician to safely use monitoring and defibrillation equipment while the HLR is operating